Copyright
©The Author(s) 2017.
World J Gastroenterol. Sep 14, 2017; 23(34): 6273-6280
Published online Sep 14, 2017. doi: 10.3748/wjg.v23.i34.6273
Published online Sep 14, 2017. doi: 10.3748/wjg.v23.i34.6273
Success (n = 19) | Failure (n = 12) | P value | |
Diameter of the first implanted SEMS, median (range), mm | 6.2 (3.1-15.9) | 6.5 (4.4-17.3) | 0.16 |
Wire passage of the first SEMS cell | 19 (100) | 8 (69.2) | 0.02 |
Diameter of wire (0.025/0.035) | 12/61 | 5/61 | 0.44 |
Catheter usage to dilate the a first SEMS cell | 8 | 7 | 0.47 |
Catheter passage of the first SEMS cell | 87.5 (7/8) | 42.9 (3/7) | 0.12 |
Dilator usage to dilate first SEMS cell | 6 | 4 | 1.00 |
Dilator passage of the first SEMS cell | 6 (100) | 1 (25) | 0.03 |
Balloon catheter usage to dilate the first SEMS lumen | 1 | 1 | 1.00 |
Balloon catheter usage to dilate the first SEMS cell | 7 | 2 | 0.42 |
Balloon catheter passage of the first SEMS cell | 7 (100) | 0 (0) | 0.03 |
The number of used dilation devices, median (range) | 1 (0-2) | 1 (0-3) | 0.76 |
Type of first SEMS used (braided/laser) | 13/6 | 8/4 | 1.00 |
Stenting order | 0.53 | ||
Left→Left | 1 | 1 | |
Left→Right | 12 | 5 | |
Right→Left | 6 | 6 | |
Procedure sessions | 0.27 | ||
1 | 16 | 12 | |
2 | 3 | 0 | |
Area of first SEMS cell, median (range), mm2 | 18.3 (3.5-18.3) | 18.3 (3.5-18.3)2 | 0.96 |
- Citation: Sugimoto M, Takagi T, Suzuki R, Konno N, Asama H, Watanabe K, Nakamura J, Kikuchi H, Waragai Y, Takasumi M, Sato Y, Hikichi T, Ohira H. Predictive factors for the failure of endoscopic stent-in-stent self-expandable metallic stent placement to treat malignant hilar biliary obstruction. World J Gastroenterol 2017; 23(34): 6273-6280
- URL: https://www.wjgnet.com/1007-9327/full/v23/i34/6273.htm
- DOI: https://dx.doi.org/10.3748/wjg.v23.i34.6273