Table 2 Set-up for the high-frequency generators for endoscopic submucosal dissection along with the IT knife for early gastric cancer
ProcedureDeviceModeOutput
ICC200
MarkingNeedle knifeForced coag20W
PrecuttingNeedle knifeENDO CUTEffect 3, 80W
Mucosal incisionIT knifeENDO CUTEffect 3, 80W
Needle knife
Submucosal dissectionIT knifeENDO CUTEffect 3, 80W
Forced coag50W
Needle knifeENDO CUTEffect 3, 80W
Forced coag50W
Endoscopic hemostasisIT knifeForced coag50W
Needle knife
Hot biopsySoft coag80W
Coagrasper
ESG100
MarkingNeedle knifeForced coag 120W
PrecuttingNeedle knifePulse cut slow40W
Mucosal incisionIT knifePulse cut slow40W
Needle knife
Submucosal dissectionIT knifePulse cut slow40W
Forced coag 250W
Needle knifePulse cut slow40W
Forced coag 250W
Endoscopic hemostasisIT knifeForced coag 250W
Needle knife
Hot biopsySoft coag80W
Coagrasper
VIO300D
MarkingNeedle knifeSwift coagEffect 2, 50W
PrecuttingNeedle knifeENDO CUT IEffect 2, CUT duration 2, CUT interval 3
Mucosal incisionIT knifeENDO CUT I or QEffect 2, CUT duration 2, CUT interval 3
DRY CUTEffect 4, 50W
Needle knifeENDO CUT IEffect 2, CUT duration 2, CUT interval 3
DRY CUTEffect 4, 50W
Submucosal dissectionIT knifeENDO CUT I or QEffect 2, CUT duration 2, CUT interval 3
DRY CUTEffect 4, 50W
Swift coagEffect 5, 50W
Needle knifeENDO CUT IEffect 2, CUT duration 2, CUT interval 3
DRY CUTEffect 4, 50W
Swift coagEffect 5, 50W
Endoscopic hemostasisIT knifeSwift coagEffect 5, 50W
Needle knife
Hot biopsySoft coagEffect 5, 80W
Coagrasper