Condition | Control (n = 15) | Test (n = 15) | P-value |
Implantation outcomes | |||
Minor bleeding | 4 | 3 | - |
Migration | 3 (20) | 2 (13) | 1.0 |
Death | 2 | 2 | - |
Number of rabbits available for retrieval | 10 | 11 | - |
Number of removable stents | 6 | 11 | 0.035 |
Stent-related complication during removal | |||
Minor bleeding | 4 | 5 | |
Hemorrhage | 1 | 0 | |
Tissue tearing | 2 | 0 | |
Death | 1 | 0 | |
Mean number of follow-up days | 25.33 | 25.85 | 0.831 |
- Citation: Liu J, Shang L, Liu JY, Qin CY. Newly designed “pieced” stent in a rabbit model of benign esophageal stricture. World J Gastroenterol 2015; 21(28): 8629-8635