Table 1 Output setting of VIO 300D for colorectal endoscopic submucosal dissection in the National Cancer Center Hospital, Tokyo, Japan
DeviceCut modeCoagulation mode
Mucosal incisionJet B knifeDry cut, [E]3 100 W
Submucosal dissectionJet B knifeDry cut, [E]3 100 WForced coag, [E]2 50 W
IT knife nanoDry cut, [E]3 100 WSwift coag, [E]2 50 W
HemostasisHemostat-YBipolar, [E]5 25 W