Analyzer and manufacturer | Assay principle | Solid phase | HCV antigens | Reaction sample volume (μL) | Time of reaction (min) | IVD registration |
Architect i2000SR, Abbott Laboratories | CMIA | Paramagnetic particles | HCr43 (Core and NS3), c100-3 (NS4A) | 20 | 20 | FDA, CE |
AxSYM, Abbott Laboratories | MEIA | Paramagnetic particles | HCr43 (Core), c200 (NS3), c100-3 (NS4A) | 33 | 30 | FDA, CE |
LiaisonXL, DiaSorin | CLIA | Paramagnetic particles | Core, NS3, NS4 | 25 | 46 | CE |
VITROS ECi, VITROS 3600, Ortho-Clinical Diagnostics | CLIA | Microwell | c22-3 (Core), c200 (NS3 and NS4), NS5 | 20 | 55 | FDA, CE |
Elecsys, Roche Diagnostics | ECLIA | Paramagnetic particles | Core, NS3, NS4 | 40 | 18 | FDA, CE |
ADVIA Centaur, Siemens | CLIA | Magnetic particles | c22-3 (Core), NS3, c200, NS5 | 10 | 58 | FDA, CE |