Copyright
©2010 Baishideng.
World J Gastroenterol. Apr 7, 2010; 16(13): 1598-1609
Published online Apr 7, 2010. doi: 10.3748/wjg.v16.i13.1598
Published online Apr 7, 2010. doi: 10.3748/wjg.v16.i13.1598
MRI platform | Avanto (Siemens, Erlangen, Germany) |
Type of pulse sequence | 3D FLASH |
TR | 2.72 ms |
TE | 1 ms |
Partition thickness | 8 mm |
Slices per slab | 10 |
Matrix | 256 × 159 |
Phase encode direction | Anterior to posterior |
Number of averages | 1 |
Sensitivity encoding factor | 2 |
Flip angle before contrast | 2º and 14º |
Flip angle after contrast | 14º |
Bandwidth | 490 Hz |
RF spoiling | Yes |
Temporal resolution | 1.98 s per slab of 10 slices. |
Precontrast scans | 10 measurements of each flip angle averaged for calculation of native T1 |
Gadolinium injection | 0.2 mmol/kg at 3 mL/s followed by 20 mL flush |
Patient respiration | Quiet breathing |
Post contrast scans | A total of 180 consecutive measurements. Inject contrast only when the 20th measurement has been completed |
Scan sections to use for processing | Center 6 image sections only |
- Citation: Thng CH, Koh TS, Collins DJ, Koh DM. Perfusion magnetic resonance imaging of the liver. World J Gastroenterol 2010; 16(13): 1598-1609
- URL: https://www.wjgnet.com/1007-9327/full/v16/i13/1598.htm
- DOI: https://dx.doi.org/10.3748/wjg.v16.i13.1598